Deluxe-100 Multi-Parameter Monitor
Applications:
Deluxe-100, with its 10.4"color display, gives a particularly clear presentationof the basic curves, optional features as RespironicsTM ETCO2 and LAN connection to central station, set it apart from much of the usual standards.Features:
•10.4 "color TFT display
•4 curves simultaneous display
•Parameters: SpO2, NIBP, Pulse, TEMP, ECG, RESP, optionalMainstream/Sidestream EtCO2 (RespironicsTM)
•Accessories for adults, children and neonates
•Visual and audible alarms with individual alarm setting
•Maximum of 480 hours trend graphs lists all parameters and trend
•Different mounting solutions: (wall, rolling stand)
•Data storage up to 400 groups of NIBP
•Network capability with central monitoring system
•Built-in lead-acid battery
Specifications:
ECG | Heart rate display range | 15bpm ~ 350bpm | |
Heart rate display accuracy | ± 1% or ± 2bpm, whichever is greater | ||
Sweeping speed | 12.5mm /s, 25mm/s, 50mm/s | ||
Sensitivity selection | ×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV | ||
CMRR | ≥ 89dB | ||
RESP | Measuring range | 0rpm~120rpm | |
Accuracy | ±5% or ±2 rpm | ||
Alarm limits setting range | 0rpm~120rpm | ||
TEMP | Measuring range | 25.0 ℃ ~45.0℃ | |
Measuring accuracy | ±0.2 ℃ | ||
Responding time | ≤150s | ||
NIBP | Measuring method | Oscillometric Technique | |
Measuring range | 0 mmHg~300mmHg | ||
Accuracy | Maximal mean difference | ±5 mmHg | |
Maximal standard deviation | 8 mmHg | ||
Measurement range | Adult | SYS 40 ~ 255 mmHg | |
MAP 20 ~ 215 mmHg | |||
DIA 10 ~ 195 mmHg | |||
Infant | SYS 40 ~ 200 mmHg | ||
MAP 20 ~ 165 mmHg | |||
DIA 10 ~ 150 mmHg | |||
Neonate | SYS 40 ~ 135 mmHg | ||
MAP 20 ~ 110 mmHg | |||
DIA 10 ~ 95 mmHg | |||
SpO2 | Measuring range | 35%~100% | |
Low perfusion capability | 0.4%~5% | ||
Accuracy | Not greater than 3% for SpO 2 range from 70% to 100% | ||
PR | Measuring range | 30bpm~240bpm | |
Measuring accuracy | ±2bpm or ±2%, whichever is greater |