UP-8000 Multi-Parameter Monitor
Applications:
UP-8000, with the CSM (Cerebral State Monitoring) module and other full-fledged functions, is especially designed for anaesthesia use.Features:
·12.1" high resolution color LCD screen
·Monitoring standard parameters of ECG, SpO2, RESP, NIBP, TEMP & PR
·CSM module to monitor parameters of CSI, EEG, EMG and BS% with alarm
·Non-invasive monitoring level of consciousness, especially for anaesthesia
·A portable handset with wireless connection to main unit for data processing
·Optimized NIBP modes for adult, pediatric & neonate
·Automatic analysis of 20 arrhythmia waveforms
·S-T segment with automatic & manual analysis
·Digital SpO2 technology with anti-motion & anti-low infusion
·Store & review 400 groups of data & 24-hour ECG waveforms
·Protection against interference from defibrillator & electrosurgical unit
·Built-in thermal printer
Specifications:
CSM | EEG sensitivity | ±400μV | |
Noise level | <2μVp-p, <0.4μV RMS, 1~250Hz | ||
CMR | >140dB | ||
Input impedance | >50Mohm | ||
Sample rate | 2000 samples/sec. (14 bits equivalent) | ||
CSI and update | 0-100. Filter 6-42Hz, 1sec. update | ||
EMG | 0-100 logarithmic. Filter 75-85Hz, 1 sec. update | ||
BS% | 0-100%. Filter 2-42Hz, 1 sec. update | ||
Wireless range | Up to 10 meters | ||
ECG | Heart rate display range | 15bpm ~ 350bpm | |
Heart rate display accuracy | ± 1% or ± 2bpm, whichever is greater | ||
Sweeping speed | 12.5mm /s, 25mm/s, 50mm/s | ||
Sensitivity selection | ×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV | ||
CMRR | ≥ 89dB | ||
RESP | Measuring range | 0rpm~120rpm | |
Accuracy | ±5% or ±2 rpm | ||
Alarm limits setting range | 0rpm~120rpm | ||
TEMP | Measuring range | 25.0 ℃ ~45.0℃ | |
Measuring accuracy | ±0.2 ℃ | ||
Responding time | ≤150s | ||
NIBP | Measuring method | Oscillometric Technique | |
Measuring range | 0 mmHg~300mmHg | ||
Accuracy | Maximal mean difference | ±5 mmHg | |
Maximal standard deviation | 8 mmHg | ||
Measurement range | Adult | SYS 40 ~ 255 mmHg | |
MAP 20 ~ 215 mmHg | |||
DIA 10 ~ 195 mmHg | |||
Infant | SYS 40 ~ 200 mmHg | ||
MAP 20 ~ 165 mmHg | |||
DIA 10 ~ 150 mmHg | |||
Neonate | SYS 40 ~ 135 mmHg | ||
MAP 20 ~ 110 mmHg | |||
DIA 10 ~ 95 mmHg | |||
SpO2 | Measuring range | 35%~100% | |
Low perfusion capability | 0.4%~5% | ||
Accuracy | Not greater than 3% for SpO 2 range from 70% to 100% | ||
PR | Measuring range | 30bpm~240bpm | |
Measuring accuracy | ±2bpm or ±2%, whichever is greater | ||
General | Power supply | 100~240VAC , 50/60Hz | |
Rechargeable Li-ion battery | 12V 2.3Ah | ||
Display mode | 12 inches TFT color LCD | ||
Alarming mode | Audible & visible alarm |