Classic-120 Multi-Parameter Monitor
Applications:
Classic-120 is a cost effetive and easy to use 12.1"patient monitor. Its extensive parameters selection allows a wide rangeof applications of this high quality model.Features:
•12.1 "color TFT display
•4 curves simultaneous display
•Parameters: SpO2, NIBP, Pulse, TEMP, ECG, RESP, optionalMainstream/Sidestream EtCO2 (RespironicsTM) & IBP (1-Channel)
•Accessories for adults, children and neonates
•Visual and audible alarms with individual alarm setting
•Maximum of 480 hours trend graphs lists all parameters and trend
•Different mounting solutions: (wall, rolling stand)
•Data storage up to 400 groups of NIBP
•Network capability with central monitoring system
Specifications:
ECG | Heart rate display range | 15bpm ~ 350bpm | |
Heart rate display accuracy | ± 1% or ± 2bpm, whichever is greater | ||
Sweeping speed | 12.5mm /s, 25mm/s, 50mm/s | ||
Sensitivity selection | ×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV | ||
CMRR | ≥ 89dB | ||
RESP | Measuring range | 0rpm~120rpm | |
Accuracy | ±5% or ±2 rpm | ||
Alarm limits setting range | 0rpm~120rpm | ||
TEMP | Measuring range | 25.0 ℃ ~45.0℃ | |
Measuring accuracy | ±0.2 ℃ | ||
Responding time | ≤150s | ||
NIBP | Measuring method | Oscillometric Technique | |
Measuring range | 0 mmHg~300mmHg | ||
Accuracy | Maximal mean difference | ±5 mmHg | |
Maximal standard deviation | 8 mmHg | ||
Measurement range | Adult | SYS 40 ~ 255 mmHg | |
MAP 20 ~ 215 mmHg | |||
DIA 10 ~ 195 mmHg | |||
Infant | SYS 40 ~ 200 mmHg | ||
MAP 20 ~ 165 mmHg | |||
DIA 10 ~ 150 mmHg | |||
Neonate | SYS 40 ~ 135 mmHg | ||
MAP 20 ~ 110 mmHg | |||
DIA 10 ~ 95 mmHg | |||
SpO2 | Measuring range | 35%~100% | |
Low perfusion capability | 0.4%~5% | ||
Accuracy | Not greater than 3% for SpO 2 range from 70% to 100% | ||
PR | Measuring range | 30bpm~240bpm | |
Measuring accuracy | ±2bpm or ±2%, whichever is greater |